Overview
HK-6539A calibration Device for Oil Conductivity Meter (Patented), in line with the GJB/J 3049-97 "light oil conductivity meter testing and calibration procedures," the use
of the requirements, mainly for light petroleum products conductivity tester testing and calibration. The device mainly consists of three parts: oil conductivity meter standard
device, special thermostatic bath and conductivity meter.
Functional characteristics
1. The oil conductivity meter standard device is precise and accurate, easy to install and use.
2. Specialized thermostatic bath with built-in compressor refrigeration, stable temperature field.
3. Adopting plane design, easy to operate the conductivity cell calibration.
Technical parameters
1. Electricity: AC 220V±10% 50Hz
2. Thermostatic bath power: 1500W
3. Conductivity meter measuring standard value range: 0 ~ 2000pS; uncertainty: ≤ 1% (k = 2)
4. Specialized thermostatic bath: temperature control range of 10 ℃ ~ 40 ℃, ± 0.01 ℃
5. Environmental requirements: temperature (22 ± 2) ℃; humidity ≤ 85%, no strong magnetic interference indoor, good grounding, no strong mechanical vibration