Overview
HK-0768 Rapid Flash Point Tester conforms to ASTM D6450 Standard Test Method for Flash Point by Continuously Closed Cup (CCCFP) Tester. The flash point tempreature is one measure of the tendency of the test specimen to form a flammable mixture with aire under controlled laboratory conditions. It is only one of a number of properties that must be considered in assessing the overall flammability hazard of a material
Features
The instrument automatically heats up, ignites, and completes the detection of the flash point. The detection speed is fast and the repeatability is good.
The instrument adopts high-voltage arc ignition, no need manual ignition, and no external gas source is required.
Only 1ml sample is required
The instrument is small in size and light in weight, suitable for portable operation
Technical parameters
Rated voltage:AC 220V±10% 50Hz
Measuring range:Ambient~180℃
Heating rate:5.5℃/min±0.5℃/min
Ignition frequency:1℃
Ambient requirements:Temperature 10~50℃;Humidity≤85%